| Product Type |
Balometer |
| Range |
Supply and exhaust: 25 to 2500 ft3/min (42 to 4250 m3/h
Air velocity: 25 to 8000 ft/min (0.125 to 40 m/s)-pitot tube
25 to 5000 ft/min (0.125 to 25 m/s)-optional probe 05949-44
25 to 2500 ft/min (0.125 to |
| Resolution |
Supply and exhaust: 1 ft3/min (1 m3/h)
Air velocity: 1 ft/min (0.01 m/s) |
| Accuracy |
Supply and exhaust: ±3% of reading ±7 ft3/min (12 m3/h)
Air velocity: ±3% of reading ±7 ft/min (0.04 m/s) |
| Temp range |
14 to 140°F (-10 to 60°C) |
| Temp resolution |
0.1°F (0.1°C) |
| Temp accuracy |
±0.5°F (0.3°C) between 32 to 160°F (0 to 71°C); ±2.0°F (1.2°C) outside of that range |
| RH range |
0 to 95%-optional probe 05949-46 |
| RH resolution |
0.10% |
| RH accuracy |
±3% |
| Display |
multi-line 6-digit LCD, 3/4" H |
| Output |
RS-232 |
| Dimensions |
Hood: 29" H x 20" dia Meter: 7 1/2"W x 4 1/2"H x 2 1/4"D |
| Brand |
Alnor |
| Model |
EBT721 |
| Power |
four rechargeable AA batteries (included) or AC adapter (included) |
| Manufacturer number |
EBT731 |